Performance Tests and Calibration 6
Keysight U1251B and U1252B User’s and Service Guide 173
10 nF 3 nF 03.000 nF 03.000 nF
10 nF 10.000 nF 10.000 nF
100 nF 10 nF 010.00 nF 010.00 nF
100 nF 100.00 nF 100.00 nF
1000 nF 100 nF 0100.0 nF 0100.0 nF
1000 nF 1000.0 nF 1000.0 nF
10 μF 10 μF 10.000 μF 10.000 μF
100 μF 100 μF 100.00 μF 100.00 μF
1000 μF 1000 μF 1000.0 μF 1000.0 μF
10 mF 10 mF 10.000 mF 10.000 mF
9
Press button to go to
mode
[b]
N/A 0 ºC 0000.0 ºC 0000.0 ºC
10 Turn the rotary swich to the
position
OPEN Input terminal open
(remove any test
leads and Short
Plugs from the
input terminal)
oPEn oPEn
500 μA 300 μA 300.00 μA 300.00 μA
5000 μA 3000 μA 3000.0 μA 3000.0 μA
11
Press button to go to
μA mode
500 μA 30 μA,1 kHz 030.00 μA 030.00 μA
300 μA,1 kHz 300.00 uA 300.00 μA
5000 μA 300 μA,1 kHz 0300.0 μA 0300.0 μA
3000 μA, 1 kHz 3000.0 μA 3000.0 μA
Table 6-4 Adjustment table (continued)
Step Test Function Cal Range Input
Cal Item
U1251B U1252B