Keysight U1251B and U1252B User’s and Service Guide 5
WARNING
Do not exceed any of the measurement limits defined in the specifications
to avoid instrument damage and the risk of electric shock.
When working above 70 VDC, 33 VAC RMS or 46.7 V peak, exercise
caution – such range pose a shock hazard.
Do not measure more than the rated voltage (as marked on the meter)
between terminals, or between terminal and earth ground.
Double-check the meter’s operation by measuring a known voltage.
For current measurement, turn off circuit power before connecting the
meter to the circuit. Always place the meter in series with the circuit.
When connecting probes, always connect the common test probe first.
When disconnecting probes, always disconnect the live test probe first.
Detach test probes from the meter before you open the battery cover.
Do not use the meter with the battery cover or part of the cover removed
or loose.
Replace the battery as soon as the low battery indicator flashes on
screen. This is to avoid false readings, which may lead to possible electric
shock or personal injury.
Do not operate the product in an explosive atmosphere or in the presence
of flammable gases or fumes, vapor, or wet environments.
Inspect the case for cracks or missing plastic. Pay extra attention to the
insulation surrounding the connectors. Do not use the meter if it is
damaged.
Inspect the test probes for damaged insulation or exposed metal, and
check for continuity. Do not use the test probe if it is damaged.
Do not use any other AC charger adaptor apart from the one certified by
Keysight with this product.