3 Instrument Operating System
Disk Drive Partitioning and Use
Backing-up the factory calibration data
To back-up the factory calibration data you will need a USB storage device.
Step Notes
1. Press System, Alignments,
Backup/Restore Align Data...
An information window appears that informs you that the
instrument must shut down to complete the backup:
"Press OK to force shut down and proceed. Press Cancel to exit."
2. Select OK.
The Alignment Data Wizard will appear:
3. Follow the on screen
instructions to backup the
calibration data to the USB drive
and save the file.
The default location is an internal flash SD
card, but you can choose another
location.