6 Keysight Connector Kit
RF Input 1, 2.4 mm adaptor installation
RF Input 1, 2.4 mm adaptor installation
Align the adaptor with the 2.4 mm input connector on the instrument and engage the
nut of the input connector with the adaptor.
Torque the bulkhead 2.4 mm input connector (with an 8 mm or 5/16" torque wrench)
to 8 inch-pounds while holding the adaptor with a 7 mm open-end wrench.
Refer to "Connector Care" on page 89.