PLZ-4W Applied Operation 6-41
• When the PLZ-4W is shipped from the factory, protection dummy plugs are
attached to the J1 and J2 connectors. To prevent the possibility of electric
shock, leave the dummy plug attached to the connectors when you are not
using them.
• To disconnect the connector, remove the lock levers located on either side
and pull the connector itself.
• Be sure to turn off the PLZ-4W when attaching or removing the connector.
• J1 and J2 connectors are physically the same. Use caution not to use the
wrong connector to prevent a malfunction.
Connecting to the connector
The connector parts needed to connect the J1/J2 connector (standard MIL connector
20-pin) are not provided. Table 6-8 shows the recommended connector.
• When using a flat cable, be sure to use a connector with a strain relief.
• To press-fit discrete wires or flat cables, be sure to use a special tool. For a
description of applicable cables and tools, see the relevant catalogs of connector
manufacturers.
Table 6-8 Connectors supported by manufacturers
An optional connector kit is available for making the connection. For details, see
page 1-8 “Analog remote contorol connector kit (OP01-PLZ-4W)”.
Manufacturer Product Notes
Omron
XG5M-2032 or XG5M-2035
XG5S-1001 (2 pcs.)
For discrete wires
Omron
XG4M-2030
XG4T-2004
For flat cables
KEL 6200-020-601 For flat cables