10. Troubleshooting
10.
Troubleshooting
The following cases may not suggest failures. Please check again before requesting
repair. When the meter does not function normally after checking the following, it
may be damaged. Remove the battery, and ask the retailer or us to repair it.
Status
Check item Measure
It does not turn on
(It does not display)
Are you pressing and holding the
Power Button for one or more
seconds?
Press and hold the Power Button
for one or more seconds.
Are of the batteries inserted
properly?
Check the signs. (
¨
P4)
Are the batteries exhausted? Replace the batteries. (
¨
P9)
Are the battery terminals dirty? Wipe them off with a dry cloth.
re you using the specied
batteries?
Check the batteries. (
¨
P4)
The LCD does not
respond
Is the screen locked? Press and hold the Menu
Button
6
to unlock the screen.
(
¨
P21)
The measurement cannot
be made
Are you using the "CL-70F
Utility"?
Stop using the "CL-70F Utility".
The measured values are
wrong
Is the Light Selection Ring in
the middle position? The light
distribution characteristics
change and suitable
measurements cannot be made.
Rotate the Light Selection Ring
until it clicks.
Is the measuring mode wrong?
(Such as measuring in Ambient
Light Mode in ash light)
Check if the measuring mode is
correct. (
¨
P22)
re you using the pre-ash
function when measuring in
Cordless Flash Mode?
In Cordless Flash Mode, the
measurement value of the main
ash may not be displayed
because the pre-ash is
measured at rst.
Cancel the pre-ash function.