3: Getting Started
22
Initial Certification
Prior to use, a qualified certifier should certify all cabinets. Under normal operating
conditions, the cabinet should be recertified at least annually and when relocated
or serviced. The certifier should perform the following tests, as recommended in
NSF/ANSI Standard Number 49 in effect when the cabinet was manufactured:
Downflow Velocity Profile Test
Inflow Velocity Test
Airflow Smoke Patterns
HEPA Filter Leak Test
Optional Canopy Alarm Test and Operation
Vibration Test *
Noise Level Test *
Lighting Intensity Test *
*These tests are user comfort related tests and may be omitted at the
user’s or certifier’s discretion.
If you have any questions regarding certification agencies or help locating one,
contact Labconco’s Product Service Department at (800) 821-5525 or +1 816-333-
8811.