7: Troubleshooting
Refer to the following table if the cabinet fails to operate properly. If the suggested
corrective actions do not solve the problem, contact Labconco for additional
assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Cabinet blower and
lights will not start
Unit not plugged
into outlet
Plug the cabinet into appropriate
electrical service
Check connection to control box on
top of cabinet
System Reset
Switch is Off
Turn on the System Reset Switch
Circuit breaker(s)
tripped
Reset circuit breakers
Keypad
disconnected or
defective
Run keypad diagnostics and check
connections
Blower will not start
Sash closed
Raise sash
Keypad
disconnected or
defective
Run keypad diagnostics and check
connections
Blower wiring is
disconnected
Inspect blower wiring
Blower motor is
defective
Replace blower motor