5: Using The Cabinet
33
Alarm Screens
Power Loss Alarm
The cabinet has lost power. Press [OK] on the Keypad to acknowledge that a
power loss occurred.
Sash Height Alarm
The sash is not in the proper operating height.
Return sash to proper working height.
Airflow Alert
The airflow patterns in the cabinet have
changed, resulting in a sudden change in the
motor speed. This is most likely due to a
blockage of the grille or the exhaust filter
outlet. It may also be caused by removal of the
work surface while the cabinet is in operation.
Remove or replace objects as necessary.
Figure 5-4b
Figure 5-4c
CHECK FOR
AIRFLOW
BLOCKAGE
Figure 5-4a