EasyManua.ls Logo

Labconco Logic+ 30348 Series - Suggested Maintenance Schedule

Labconco Logic+ 30348 Series
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6: Maintaining The Cabinet
48
Suggested Maintenance Schedule
Maintenance Frequency
Activity
Weekly
Monthly
Annually
Disinfect interior surfaces (with
suitable chemical disinfectant)
Wipe down interior surfaces after
contact time elapsed with 70%
alcohol solution
Clean sash glass and UV lamp with
suitable glass cleaner
Operate cabinet blower, noting
Filter Life percentage in log
Using 70% alcohol solution, wipe
down cabinet exterior
Disinfect and lift work surface.
Surface disinfect the area beneath
with suitable chemical disinfectant
Wipe down area beneath work
surface after contact time elapsed
with 70% alcohol solution
Check all service valves (if
equipped) for proper operation
Check the UV Lamp Hourmeter,
noting in log
Check prefilter on top of unit;
replace if necessary
Have the cabinet re-certified by a
qualified technician

Table of Contents