7: Troubleshooting
58
PROBLEM
CAUSE
CORRECTIVE ACTION
Contamination of
work in the
cabinet
Improper technique
or procedure for the
cabinet
See Working In the Cabinet section
at the end of Section 5 in this
manual
Restriction of the
return air slots or
grille blockage of
the exhaust outlet
Ensure that all return air slots,
grilles and the exhaust outlet are
unobstructed
External factors are
disrupting the
cabinet airflow
patterns or acting
as a source of
contamination
See Working In the Cabinet section
at the end of Section 5 in this
manual
Cabinet is out of
adjustment/HEPA
filter(s) are
defective
Have cabinet recertified. If HEPA
filters are defective or damaged,
replace them