Proline Kryomats
100 Maintenance YAUE0007 / 2/26/2009
9.3.4.2.2 Decalcifying the water cooling circuit
At regular intervals of 3 months or longer, the water-cooled condenser must be decalcified or cleaned.
This depends on the hardness of the cooling water and the degree of soiling. Drain according to (Ö
9.2.2).
Required
equipment:
Two containers of approx. 10 to 20 liters volume.
Use a suitable pump (drum pump) or a hose with funnel. Place the funnel as high as
possible so that the device can fill quickly.
Fit connecting hoses between container, pump, cooling water inlet and between cooling
water outlet and back to container.
Change containers when
top one is empty.
Acting time:
Continue the pump stage until most of the foamy reaction, usually at the
start, has decayed. Generally, this is achieved after about 15 to 30 minutes.
Decalcifier:
Water with LAUDA Decalcifier LZB 126.
It is essential to follow the safety instructions when handling the chemicals.
Flushing:
Allow at least 30 liters of water to flow through.
9.3.5 Repair information
If you need to send in a unit for repair, it is essential to first contact the LAUDA Service Constant
Temperature Equipment Ö 9.3.7.
)
When sending in the unit, ensure that it is carefully and properly packed. LAUDA
cannot be held liable for any damage caused by improper packing.
Standard & Alternative

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lauda RP 3090 C and is the answer not in the manual?

Lauda RP 3090 C Specifications

General IconGeneral
BrandLauda
ModelRP 3090 C
CategoryThermostat
LanguageEnglish

Related product manuals