YAUE0007 / 2/26/2009 Starting up 41
7.5 Important settings
7.5.1 Temperature setpoint setting
The setpoint is the temperature, which the thermostat should reach and maintain constant.
Master (main level)
RDr
Press key until
RDr (Setpoint) appears.
Press key, display flashes.
or
Enter the setpoint with the two keys (Ö 7.4.1 General key functions
and pilot lamps).
Wait 4 seconds or
Display flashes 4 seconds Æ new value is automatically accepted,
or
value is accepted immediately with Enter key.
)
For safety reasons the setpoint can only be set up to 2°C above
upper limit of the operating temperature range for the relevant
device type.
In the following cases, the manual setpoint entry is blocked:
Setpoint is taken from the analog module, from the programmer in
the Command Console or via the serial interface.
When the setpoint temperature is to be lowered, it may take up to
one minute before the blue LED
lights.