Proline Kryomats
54 Starting up YAUE0007 / 2/26/2009
7.6.4 Restoring works settings
All works settings, apart from the control parameters and the sensor calibration, are restored.
Command
− Works settings
− Open the device parameter menu via
the soft key Menu .
− With the cursor keys continue to
Æ Settings Æ Works settings .
− The window shown opposite appears.
− Master and then only control par.
int. is shown as a possible choice.
There are however various possibilities,
which can be selected with or :
− Under All modules Master, Command
and all connected modules are reset to
the works setting with all default .
− Under Master you have the choice
between:
− all default , then all Master settings
are reset,
− only control para. int. for the
internal control parameters,
− only control para. ext. similar for
external,
− only miscellaneous which resets
setpoint, pump level, maximum
current consumption, control to
internal and auto start to "Auto".
− Under Command all command settings
are reset with All default .
− Confirm selection with .
− Confirm the control dialog shown on the
left with or cancel with .
− Return to measurement window with
End or .
Hel
Menu End T
set
T
fix
All modules
Master
Command
Cool
all default
only control par. int
only control par. ext
only miscellaneous
Hel
Menu End T
set
T
fix
Confirm input!
Enter key: Continue
Escape key: Cancel