Proline Kryomats
106 Technical data YAUE0007 / 2/26/2009
11 Technical data and diagrams
The figures have been determined according to DIN 12876.
RP 3050 C RP 3050 CW RP 4050 C RP 4050 CW
Operating temp.- ACC range °C -50...200
Ambient temp. range °C 5...40
Relative humidity maximum relative humidity 80 % for temperatures up to 31 °C, decreasing linearly to 50 % relative
humidity at 40 °C
Temperature range for
storage
°C -20…44
at water-cooled devices the condenser must be completely empty (Ö 9.2.2)
Setting resolution °C 0.1 / 0.01 (Master); 0.01 (Command)
Display resolution °C Master: 0.01 Command: 0,1 / 0,01 / 0,001
Display accuracy ±0.2 °C can be calibrated additively (Æ Section 1.2 last Point)
Temperature stability ±K 0.05
Safety equipment Class III, FL suitable for flammable and non-flammable liquids
Maximum cooling water
consumption: temperature
15 °C, pressure 3 bar ~
[L/h] 700 700
Water-cooling connections
(DIN EN 10226-1)
G ¾” G ¾”
Heater power
400 V; 50 Hz
kW maximum 3.5
with heat
transfer oil
200 °C kW Data coming soon Data coming soon 5.00 7.00
with ethanol 20 °C kW 5.00 6.00
0 °C kW 3.0 3.5
-20 °C kW 1.6 1.8
-30 °C kW 1.0 1.1
-40 °C kW 0.5 0.6
-50 °C kW 0.25 0.25
-60 °C kW
-70 °C kW
-80 °C kW
Cooling power at 20°C t
amb
@ bath temp. (Pump Level
-90 °C kW
Pump type Pressure pump, 4 power levels (level 5 to 8)
Discharge pressure max. bar 0.5 at pump power level 8
Flow rate max. (pressure) L/min 19 at pump power level 8
Hose connections Thread M16x1; olives 13 mm external diameter
Bath volume from...to L 23...31 32...44
Bath opening B x L mm 350 x 200 350 x 350
Bath depth / usable depth mm 250
Height to top of bath mm 905
Overall dims. B x L mm 600 x 700