EasyManuals Logo

Lauda RP 3090 C User Manual

Default Icon
115 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #46 background imageLoading...
Page #46 background image
Proline Kryomats
46 Starting up YAUE0007 / 2/26/2009
7.5.4 Activate external pump
As an option an external pump is available for the Proline Kryomats for external applications. This pump
can be set on / off manually in the menu shown below. Another possibility is the automatic mode. In this
case the pump is switched according to the unit status standby / running.
Command
Ext. Pump
Open the device parameter menu via the
soft key Menu .
Change from Pump Æ Ext. pump
using .
With or you enter the illustrated
window.
Use or to switch the pump off,
permanently on or to set the automatic
mode. Confirm your selection with or
End ,
or quit the window with without making
any changes.
Automatic mode means that the external
pump is switched off while the unit is in
standby mode or it is switched on while
the unit is running.
7.5.5 Activating external control
An external temperature probe can be connected to the Proline Thermostats. How this is done is
explained in Section 7.5.2. If the bath temperature is to be controlled using this sensor instead
of the
internal sensor, the setting can be made here.
Furthermore, control can also occur based on the signal from the analog or serial module (Ö 4.8).
Master
Bml
and 2 x
Call the source selection for the control Bml .
The momentary setting for the source is displayed,
here
0 for internal, because control takes place using the
temperature signal from the internal temperature probe.
The source display flashes.
Bml 0
°C
Hel
p
Menu End T
set
T
fix
off
on
auto.
Ext. pump

Table of Contents