Proline Kryomats
YAUE0007 / 2/26/2009 Starting up 45
Wait 4 seconds or
Display flashes 4 s Æ new value is automatically accepted, or
value is immediately accepted with Enter key.
Command
Pump Level
Open the device parameter menu via the
soft key Menu .
Change from Pump Æ Pump Level
using .
With or you enter the illustrated
window. Level 5 is active.
Select another pump level with or
and confirm with or End ,
or quit the window with without making
any changes.
It is not possible to select the pump levels
1 to 4.
Standby activation
+
Standby activation
(Pump, heater and refrigerating machine
are deactivated when the green LED in
the lower part of the key is lit).
Please exercise caution when thermostat is in standby mode.
The following settings/ actions may start the thermostat unintentionally from the standby
mode:
a previously activated timer mode (Ö 7.10), because a started timer continues
to run!
“Start” command via interface
s (Ö 8).
Pump Level
Hel
p
Menu End T
set
T
fix
Level 8
Level 7
Level 6
Level 5
Level 4
Level 3
Level 2
Level 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lauda RP 3090 C and is the answer not in the manual?

Lauda RP 3090 C Specifications

General IconGeneral
BrandLauda
ModelRP 3090 C
CategoryThermostat
LanguageEnglish

Related product manuals