Proline Kryomats
56 Starting up YAUE0007 / 2/26/2009
− The temperature measurement device
shows the true temperature value (with
glass thermometers consider the
correction where applicable!).
− Change the display in the adjacent
window to the true value with cursor or
soft keys and accept with or End ,
− or quit the window with without making
changes.
7.6.7 Restoring the works setting of the internal temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Command
− Default
− Open the device parameter menu via
the soft key Menu .
− With the cursor keys continue to
Æ Settings Æ Calibration Æ
intern Pt100 Æ Default .
− The window shown adjacent appears.
− Confirm selection with .
− Confirm the control dialog on the right
with or cancel with .
− Return to the measurement window with
End or .
Hel
Menu Screen T
set
T
fix
6 7 8 9 0
1 2 3 4 5
Temperature value of the ref.
temp. measurement device:
20.15
Min: -50.0°C Max: 302.00
Hel
Menu End T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Hel
Menu End T
set
T
fix
Confirm input!
Enter key: Continue
Escape key: Cancel