Proline Kryomats
64 Starting up YAUE0007 / 2/26/2009
7.8.4 Creating or modifying a program (Edit)
Here, there are the following functions:
Entry of a program.
Display of the program data of a saved program and modification of the segment data.
Insertion or appending of a new segment.
Deletion of a segment.
)
In addition, when a program has just been executed, new segments can be inserted
and existing ones modified, even the currently active segment. Furthermore, all
segments, except the currently active one, can be deleted at any time.
Modifications to the currently running segment are possible. The segment then
continues as though the modification had been applicable since the start of the
segment.
However: If the new segment time is shorter than the segment time that has already
run, then the program skips to the next segment.
If a segment time >999h: 59min is required, then this time period must be shared
over a number of consecutive segments.
Entering a program:
Program example (Ö 7.8.1)
Command
Programmer Program1 Edit
Modify
In the Edit menu one can Modify or
Delete a program.
Press the key.
Continue to Modify . with the key .
There is the possibility of modifying single
segments, i.e. segments can be entered
as new, changed and also deleted.
Hel
p
Menu End T
set
T
fix
Status
Edit
Loops
Graph
Info
Modify
Delete