Proline Kryomats
108 Technical data YAUE0007 / 2/26/2009
RP 3090 C RP 3090 CW RP 4090 C RP 4090 CW
Operating temp.- ACC range °C -90...200
Ambient temp. range °C 5...40
Relative humidity maximum relative humidity 80 % for temperatures up to 31 °C, decreasing linearly to 50 % relative
humidity at 40 °C
Temperature range for
storage
°C -20…44
at water-cooled devices the condenser must be completely empty (Ö 9.2.2)
Setting resolution °C 0.1 / 0.01 (Master); 0.01 (Command)
Display resolution °C Master: 0.01 Command: 0,1 / 0,01 / 0,001
Display accuracy ±0.2 °C can be calibrated additively (Æ Section 1.2 last Point)
Temperature stability ±K 0.05
Safety equipment Class III, FL suitable for flammable and non-flammable liquids
Maximum cooling water
consumption: temperature
15 °C, pressure 3 bar ~
[L/h] 700 --- 800
Water-cooling connections
(DIN EN 10226-1)
G ¾” G ¾”
Heater power
400 V; 50 Hz
kW maximum 3.5
with heat
transfer oil
200 °C kW 3.0 4.0 3.00 4.00
with ethanol 20 °C kW 3.0 4.0 3.00 4.00
0 °C kW 2.9 3.7 2.90 3.70
-20 °C kW 2.5 3.1 2.5 3.1
-30 °C kW 2.3 2.7 2.3 2.7
-40 °C kW 2.0 2.3 2.0 2.3
-50 °C kW 1.6 1.8 1.6 1.8
-60 °C kW 1.3 1.4 1.3 1.4
-70 °C kW 0.8 0.9 0.8 0.9
-80 °C kW 0.5 0.5 0.5 0.5
Cooling power at 20°C t
amb
@ bath temp. (Pump Level
-90 °C kW 0.15 0.15 0.15 0.15
Pump type Pressure pump, 4 power levels (level 5 to 8)
Discharge pressure max. bar 0.5 at pump power level 8
Flow rate max. (pressure) L/min 19 at pump power level 8
Hose connections Thread M16x1; olives 13 mm external diameter
Bath volume from...to L 23...31 32...44
Bath opening B x L mm 350 x 200 350 x 350
Bath depth / usable depth mm 250
Height to top of bath mm 905
Overall dims. B x L mm 600 x 700
Overall dim. H mm 1160