Aux. high-level DC Inputs
Num ber of Aux channels 4 on Aux. connector
ADC Resolution 16 bits
Sam pling
1 to 256Hz or 25 t o 200Hz rat e, sim ult aneous sam pling all
channels
I nput im pedance 47 Kohm s
Hardware gain 1 ± 2%
Max input before clipping ± 2.5 V
Bandwidt h
DC to 70Hz ( 6dB)
Modes of operation
I m pedance
< 2 to > 90 KOhm , m easured with 0.075uA pp, 5Hz signal. Measurem ent
accuracy ± 15% .
Calibrat ion
0.5m V pp square wave @ 1Hz at ADC input. Am plit ude accuracy ± 5% .
Connections, ports and controls
Patient Connect ion Unit
(36 pin MDR connector)
Connector for t ouch- proof Pat ient Connect ion Unit (PCU), ei-
ther Leaded or Clickon
Front - panel push- but tons 1 on/ off push- but ton Pat ient Event during r ecord
Aux. Connect or 1
( RJ45 connect or)
USB port ( isolat ed from pat ient )
Aux. Connect or 2
(15 pin Micro- D connector)
Ext ernal Pat ient Event input
Bluetoot h Module I / O
Nonin XPOD SaO2 probe input
4 Aux. High-level DC input s
CF card port 1 TypeI I socket ( 3.3V)
Bat t eries 3 type PP3/ MN1604
1 type BP511 Lit hium rechargeable
I nt ernal beeper sounds when:
door open
bat tery low
car d st orage low
Pat ient Event
Back-light display
Current Tim e and Dat e
Recording Tim e
Bat t ery Life rem aining
Disk Storage rem aining
Door open warning
Recording format
Native European Dat a Form at ( EDF)
Native MSDOS/ Windows card dat a st ruct ure
Physical characteristics
Weight 500g including card and 3 PP3 bat t eries
450g including card and Li rechargeable batt ery
Size
16cm x 8cm x 3.5cm