7 Parameters
43
P Units: Atm
Bar
kPa
m H2O
p
si
mm H
g
Pressure Display Units. Available when the
pressure compensation is set to “Manual”
Ø
ØØ
Ø ×
××
×
1.0
1.01.0
1.0
Bar
In
p
ut Pressure
Fixed pressure level when
compensation is manual
Ø
ØØ
Ø ×
××
×
Simulate In
p
ut
Press “ENTER” to Select the Input
Simulation Mode
1.0
1.01.0
1.0
ppm
Simulate : 12.00mA
Simulate sensor input. Use the “Up”
and “Down” keys to scan through the
input range to test the Set Point
Relays and Current Outputs
UNITS
The MOD53 primary display can be setup to display in % (saturation), ppm
(concentration), pO2 (partial pressure of Oxygen) or input current. This is achieved
by simply setting the appropriate units. The relationship between these three
parameters is determined by several factors including temperature, pressure and
the salinity of the solution being measured. (See Appendix A).