Appendix C – Factory Default Setups
89
Parameters Galvanic Polargraphic
Current Outputs
In
p
ut A Sensor Sensor
Out
p
ut A 4-20mA 4-20mA
Out
p
ut A Zero 0.0% 0.0%
Out
p
ut A S
p
an 100.0% 100.0%
In
p
ut B Tem
p
erature Tem
p
erature
Out
p
ut B 4-20mA 4-20mA
Out
p
ut B Zero 0.0°C 0.0°C
Out
p
ut B S
p
an 100°C 100°C
Pressure Input
Mode In
p
ut In
p
ut
Units Atm Atm
4mA Level 0.40 Atm 0.40 Atm
20mA Level 2.00 Atm 2.00 Atm
Pressure Dam
p
in
g
Disabled Disabled
Pressure Limit A
(
“From”
)
0.00 Atm 0.00 Atm
Pressure Limit B
(
“To”
)
0.00 Atm 0.00 Atm
Calibration
Mode On-Line On-Line
Calibration Access No No
Configuration
Lan
g
ua
g
e En
g
lish En
g
lish
T In
p
ut LTH 1K* BJ 22K**
Bias Volta
g
e 0.000V +0.675V**
Membrane Correction Factor 3965* 2220**
Errors Enabled Enabled
In
p
ut Filter Out Out
* Default for use with the OE15 Probe
** Default for use with the ProcessProbe