Making measurements - a tutorial Chapter 4
Mastersizer 2000 Page 4-9
Select OK.
5. In the Measurement Display window press the Next button to add the sam-
ple. The display will change. Add one drop of the cream to the tank and wait
for it to diffuse through the system. Add enough drops to make the blue Laser
Obscuration bar move into the green area (try and achieve a mid value).
6. Once satisfied that enough sample has been added, press the Start button. The
sample is measured and the result (size distribution) calculated. The measure-
ment record is automatically saved to the records buffer.
7. At the end of the measurement a dialogue appears, asking if the user wants to
make another measurement. Select Finished measuring. The Measurement
Display disappears and the software returns to the main window.
8. The measurement is now complete. Click on the Records tab, select the
record and then select one of the report tabs to view the results in detail.