Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 5-3
5
ALARM SILENCE
Audible alarms may be silenced, while visual alarms remain active. The alarm silence
function is controlled by pressing the Alarm Silence Button.
The Alarm Bell provides visual feedback when the Rad-8 audible alarms are silenced.
Alarm Silence function when monitoring a patient:
Power-On – Alarms active, Alarm Bell flashes red for high priority alarms
Press Once – Alarm suspended for 120 seconds. Audible alarms are silenced. The Alarm
Bell flashes red for high priority alarms.
Press Twice - Returns to audible alarms active. Audible alarms return,
Repeated pressing of the Alarm Silence Button will cycle through alarm silence options above.
Alarm Silence function when not monitoring a patient:
Power-On – Alarms active, Alarm Bell is not flashing.
Press Once – Device is silenced until it is cycled off/on or until monitoring begins. The
Alarm Bell is not flashing
ALARM BELL
The Alarm Bell flashes red for high priority alarms. Pressing the Alarm Silence Button once
silences the audible alarm for 120 seconds (default) while the Alarm Bell flashes to indicate
an alarm condition. If the high priority alarm condition is resolved during the Alarm silence
interval, the Alarm Bell stops flashing. If the high priority alarm condition remains (Alarm
Bell flashing red), pressing the Alarm Silence button again activates the audible alarms and
the Alarm Bell continues to flash red. The Alarm Bell stops flashing when the high priority
alarm conditions are resolved.
ALARM ON/OFF
Alarm OFF: All audible alarms are muted and visual alarms are active. Audible alarms
are restored when power is turned off and back on.
Alarm Off (re): Audible alarms are muted, but the alarm “beeps” twice every three minutes
to remind the user that the Rad-8 is currently in an alarm status but that the audible alarm
is muted. Visual alarms are active in this mode. If the alarm is violated, the associated
parameter/measurement label and value flash, and the alarm bell flashes red for high
priority alarms. Audible alarms are restored when power is turned off and back on.
messages

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo Rad-8 and is the answer not in the manual?

Masimo Rad-8 Specifications

General IconGeneral
DisplayLED
Pulse Rate Range25-240 bpm
Pulse Rate Accuracy±3 bpm
Trend Time72 hours
SpO2 Range0-100%
SpO2 Accuracy± 2 digits (70-100% SpO2)
Power SupplyAC adapter or rechargeable battery
Perfusion Index (PI) Range0.02% - 20%
PVI Range0-100
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-4°F to 140°F (-20°C to 60°C)
Relative Humidity10-95% non-condensing
AltitudeUp to 15, 000 ft (4, 572 m)