EasyManua.ls Logo

Masimo Stork Vitals+ - Temperature Specifications; Display Range and Display Resolution; Accuracy; Electrical

Masimo Stork Vitals+
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Stork Vitals+ Appendix
Masimo Corporation 47 www.masimo.com
The below Bland-Altman plot represents the distribution of points in the difference between
the RD SET Adhesive Sensor and the reference SaO
2
(SpO
2
SaO
2
) versus the reference
(SaO
2
) under no motion with an upper 95% and lower 95% limits of agreement.
Figure 2: RD SET Adhesive Sensor (70-100%)
Temperature Specifications
Display Range and Display Resolution
Measurement
Display Range Resolution
Temperature 25°C to 43°C (77°F to 109.4°F)
0.1 °C (0.1 °F)
Stork provides a surface temperature of the skin where the Stork Sensor is applied. The
displayed temperature may require time to equilibrate after initial application (up to 7
minutes).
Accuracy
Temperature
Laboratory
25°C to 43°C (77°F to 109.4°F)
±0.3°C (±0.54 °F)
Application Site Foot
Electrical
Battery - Stork Sensor
Type Li-ion Rechargeable
Run Time 16 hours [5]
Charging Time
2 Hours [6]

Table of Contents

Related product manuals