Stork Vitals+ Appendix
Masimo Corporation 56 www.masimo.com
Symbol Description Symbol Description
Keep dry
Do not use if package is damaged
Storage humidity limitation
Wireless Symbol level
Atmospheric pressure limitation
Greater Than
Less Than - --
Instructions/Directions for Use/Manuals are available in electronic format
@http://www.Masimo.com/TechDocs
Note: eIFU is not available in all countries.
Citations
[1] The specifications reflect the representative accuracy of the RD SET Adhesive sensor using Masimo SET
Technology that has been validated for no motion accuracy in human blood studies on healthy adult male and
female volunteers with light to dark pigmented skin in induced hypoxia studies in the range of 70%-100% SpO
2
against a laboratory co-oximeter
[2] The specifications reflect the representative accuracy of the RD SET Adhesive sensor using Masimo SET
Technology has been validated for motion accuracy in human blood studies on healthy adult male and female
volunteers with light to dark pigmented skin in induced hypoxia studies while performing rubbing and tapping
motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an
amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70%-100% SpO
2
against a laboratory co-
oximeter
[3] The Masimo Stork has been validated for low perfusion accuracy in bench top testing against a simulator
with signal strengths as low as 0.02% for the simulated saturation range of 70% to 100%.
[4] The Masimo Stork has been validated for pulse rate accuracy for the range of 25-240 bpm in bench top
testing against a simulator and Masimo’s simulator with signal strengths of greater than 0.02%.
[5] This represents the approximate run time when it is continuously operating with the Bluetooth active
after a fully charged battery.
[6] The battery recharge time shall be no longer than 2 hours to reach 80% charge capacity at operating
temperature of 25°C (77°F) ambient temperature and might not charge completely under elevated ambient
temperature.