Stork Vitals+ Appendix
Masimo Corporation 57 www.masimo.com
Service and Maintenance
Cleaning
The Stork Sensor, the Stork Boot and the straps should be cleaned daily or before each use
or when they appear soiled or dirty.
WARNING: Before cleaning, remove the Stork Sensor from the Stork Boot. Do not put the
sensor in the washing machine.
CAUTION: Do not clean the sensor with undiluted bleach, petroleum-based products,
acetone, or other harsh solvents. Clean only with the solutions specified to prevent damage
to the device.
Dampen a cloth with water and mild detergent (dish soap) to wipe the
sensor or boot of the Stork System (sensor is shown). Disinfectant wipes
or 70% isopropyl alcohol (IPA) can also be used.
Air dry when cleaning is done.
You can also place the boot (and the straps if desired) in the washing machine. It is
recommended to place the boot and straps into a delicates bag and wash with the baby's
clothes.
WARNING: Ensure the baby's foot and the Stork Boot are completely dry before putting
the boot and sensor back on the baby's foot.
Note: The Stork Camera does not require cleaning with the exception of wiping the surface
with a damp cloth.
Customer Support
For product support, along with troubleshooting for your Stork product, please go to the
Stork Support page www.masimostork.com/en-us/support/contact-us.html.