Stork Vitals+ Appendix
Masimo Corporation 49 www.masimo.com
Electromagnetic Emissions
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Complies
Electromagnetic Immunity
Electromagnetic Immunity
The ME Equipment is intended for use in the electromagnetic environment specified below. The customer or
the user of the ME Equipment should assure that it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance Level Electromagnetic Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 8 kV
contact
+/- 15 kV air
+/- 8 kV contact
+/- 15 air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/ burst
IEC 61000-4-4
+/- 1 kV for
input/ output
lines
+/- 1 kV for
input/ output
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
+/-1 kV line(s)
to line(s)
+/-1 kV line(s) to
line(s)
Mains power quality should be that of a typical
commercial or hospital environment.
Conducted RF
IEC 61000-4-6
3 Vrms 3 Vrms Performed over 0.15-80 MHz
6 Vrms 6 Vrms
Performed on the following ISM (industrial, scientific
and medical) bands of frequency: The bands between
0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz;
13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz;
and 40,66 MHz to 40,70 MHz. The amateur radio bands
between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz,
3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3
MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07
MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to
24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0
MHz
Power frequency
(50 / 60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m Power frequency magnetic fields should be at levels
characteristic of typical location in a typical hospital
environment.
Voltage dips on
power supply
input lines
IEC 61000-4-11
0% UT
1
, 0.5
cycle, at 0°,
45°, 90°, 135°,
180°, 225°,
270°, and
315°;
0% UT 1 cycle,
and 70% UT
25/30 cycles at
0% UT
1
, 0.5 cycle,
at 0°, 45°, 90°,
135°, 180°, 225°,
270°, and 315°;
0% UT 1 cycle,
and 70% UT 25/30
cycles at 0°
Mains power quality should be that of a typical
commercial or hospital environment.