2
WARNING: Reusable accessories that
must be forwarded to an Authorized
Mectron Service Center.
All the parts must be cleaned and sterilized
in accordance to the procedures described
in this Manual, before they are sent to
an Authorized Mectron Service Center.
Incorrectly treated and bio-contaminated
parts will not be accepted by the Authorized
Mectron Service Centers.
2 CLEAN/FLUSH FUNCTION
The CLEAN/FLUSH function (according to the device used) enables to run a cleaning cycle on the
irrigation circuit of the handpiece used during the treatment, by following the steps described
in this chapter.
CAUTION: Clean/ush function. The
CLEAN/FLUSH function must be used after
every treatment, before starting the cleaning
and sterilization procedures.
CAUTION: If the tubes are not cleaned,
the crystallization of the salts may seriously
damage the device.
CAUTION: The handpiece and the cord
cannot be separated.
2.1 Preparation
Close the ow of the adjuster.
Disconnect the ow adjuster from the
irrigation bag and from the peristaltic
pump tube or from the tube of the bone
grafting kit, if used.
Dispose of the ow adjuster correctly,
following the norms in force locally on
the disposal of hospital wastes;
1
Prepare an enzymatic detergent solution
with pH 6-9, as per the instructions of the
manufacturer;
2