4.2.3.8 Irrigation Bag Support Rod and Fixed Handpiece Support
Clean the surfaces of the rod and of the
handpiece holder with a soft cloth with
low fiber-release, dampened with an
enzymatic detergent solution;
26
Remove the residues of detergent
with a soft cloth with low fiber-release
dampened with distilled water;
27
4.2.4 Automatic Cleaning
4.2.4.1 Materials Required
Alkaline detergent: neodisher® FA (0.2
% v/v);
Neutralising liquid: neodisher® Z (0.1
% v/v);
Thermal disinfector;
Metal tray;
Adaptors.
NOTE: Place the accessories in the washing
machine so that dead zones do not arise and
the water can properly drain. Also, make sure
that the reusable inserts are properly held in
place in the washing basket and cannot move
during the washing process, as shocks could
damage them.
WARNING: Avoid overloading. Failure to
comply overloading compromises cleaning
eectiveness.
WARNING: Once the cleaning cycle in
thermodisinfector has been completed, the
handpiece keeps the washing temperature
for a long time. When removing the
handpiece from the thermodisinfector take
proper measures to avoid damages for the
user.
CAUTION: Due to its shape, the
handpiece can roll. For this reason, when it
is not in use, it shall always be placed on its
support.
CAUTION: Before starting to clean the
thermal disinfector, assess the entity of
dirt and, if necessary, perform a pre-wash
(
Chapter 4.2.2 on page 13
) to avoid
repeating the entire cleaning cycle.