CLEANING AND STERILIZATION
4.2.5 Cleaning Check
Once you’ve nished the cleaning
operations, check all the objects under an
adequate source of light, if need be also
with a magnifying glass, paying attention
to the details that could conceal dirt
residues (threading, cavities, grooves)
and, if need be, repeat the chosen
cleaning cycle. Finally, check the integrity
of those parts and elements that may
have worn out during their use;
1
4.2.6 Drying and Lubrication
Thoroughly dry all parts of the handpiece
with cord, especially the electrical
contacts, blowing on them with
compressed air;
1
CAUTION: The electrical contacts
of the handpiece cord connector must
be dry, before and after the sterilization
cycle, before connecting the cord to
the device. Always make sure that
the connector electrical contacts are
perfectly dry; if need be, dry them by
blowing compressed air.