4
To make the Clean function start while
the PUMP/CLEAN writing is ashing,
press the foot pedal once and release
it: PUMP/CLEAN stops ashing and the
cleaning cycle starts;
NOTE: The Clean function can be
interrupted at any time by pressing
the PUMP/CLEAN writing again or,
alternatively, by pressing on foot pedal.
The keyboard is enabled again, and
displays the last setting used;
2
As soon as the peristaltic pump starts, the
entire scale of values of the “irrigation”
section lights up and during the liquid
passage, the value of the irrigation shifts
from 6 a 0 . The cycle lasts 20 seconds.
Once it has ended, the keyboard is
enabled again, and displays the last
setting used;
Proceed with the procedure described in
Chapter 2.4 on page 5
.
3
2.3 PIEZOSURGERY
®
white
To enter the cleaning mode keep the
PUMP/FLUSH button on the touch
keyboard pressed; the LED of the level 6
of the function Irrigation ashes; all other
buttons are disabled. To stop the FLUSH
function release the button;
Proceed with the procedure described in
Chapter 2.4 on page 5.
1
1
0