16
Using a 20 ml disposable syringe, inject
20 ml of cold demineralised water 3 times
inside the cavity of the insert;
9
x3
4.2.2.5 Inserts Torque Wrench
Soak the wrench in the enzymatic
detergent solution
e)
for 10 minutes at 40
°C ±2 °C;
e) Process validated by independent body with
enzymatic detergent Enzymec (Mectron).
10
Using a 20 ml disposable syringe, ush
the surface of the wrench three times with
20 ml of enzymatic detergent solution;
11
Delicately brush the surface of the
wrench under running water (drinking
water, temperature 20 °C+/- 2 °C), in
particular the parts that are dicult to
access (cavities, structured surfaces, etc.)
using a brush with soft nylon bristles;
The wrench must be brushed until no
residues can be seen (at least 30 seconds);
12