14
4.2.2.3 Handpiece with Cord
Clean the surface of the scaler handpiece
and of its connector with a clean, soft
cloth with low ber-release, dampened
with a detergent solution (pH 6-9)
c)
;
c) Process validated by independent body with
enzymatic detergent Enzymec (Mectron).
2
Delicately clean the surface of the
handpiece using the enzymatic detergent
solution with the aid of a brush with soft
nylon bristles, placing special care to the
following areas:
• threading of the handpiece;
• titanium stem;
3
• front terminal in its external and
internal parts.
Completely rinse with tap water to
eliminate any detergent residues and
perform the last rinse with distilled water;
4