9
CLEANING AND STERILIZATION
4 CLEANING AND STERILIZATION
This table is purely indicative.
For the complete cleaning and sterilization
procedures of the individual parts, refer to the
chapters indicated in the table.
CAUTION: Methods not included in the
table below must not be used.
Device surface and Pedal
Step Ref. Procedure
I
4.1
Manual cleaning with detergent
Accessories (Handpiece, Inserts, Torque Wrench, Peristaltic Pump Tubes, Handpiece
Mobile Support, Tube-to-Tube Connector, Irrigation Bag Support Rod and Handpiece
Fixed Support)
Step Ref. Procedure
Handpiece /
Rod / Hand-
piece Fixed
Support
Pump Tubes
/ Handpiece
Mobile Sup-
port
In-
serts
Tube-
to-Tube
Connec-
tor
Torque
Wrench
II
4.2.2
Pre-Cleaning
a)
Manual clean-
ing with enzy-
matic deter-
gent
Immersion into enzymatic detergent
III
4.2.3
Manual clean-
ing
Manual clean-
ing with enzy-
matic deter-
gent
Immersion
into enzy-
matic deter-
gent
Immersion into enzymatic de-
tergent and ultrasonic wash-
ing
4.2.4
Automatic
cleaning
Thermal disinfector with detergent
IV
4.2.5
Cleaning
Check
Visual inspection with 2,5x lens
V
4.2.6
Drying and lu-
brication
Drying
Drying
and Lu-
brication
VI
4.2.7
Sterilisation Packaging, Sterilisation and Storage
a) Optional.