22
Leave to soak for 10 minutes at 40 °C ±2
°C. This procedure reduces the quantity
of blood, protein, and mucous present in
the insert;
6
During immersion into the enzymatic
solution, delicately brush the surface of
the insert with a brush with soft nylon
bristles to eliminate all traces of visible
dirt. Accurately clean the dicult areas
such as sharp edges and especially the
interstices between the cutting cusps;
7
Delicately brush the surface of the insert
again with the brush with soft nylon
bristles;
8
Using a syringe, inject the enzymatic
detergent solution 3 times inside the
cavity of the insert to eectively remove
the residues from the internal surface;
9
x3