23
CLEANING AND STERILIZATION
Rinse the internal canal of the insert with
distilled water injected under pressure
(3,8 bar) for at least 10 seconds to
eliminate all residues;
10
3,8 bar
10 sec.
Place the insert in the ultrasound tank
submerged by the enzymatic detergent
solution at 40 °C ±2 °C, for at least 10
minutes;
11
Remove the insert from the ultrasonic
tank and rinse with distilled water;
12
Delicately brush the surface of the insert
again with the brush with soft nylon
bristles;
13