24
Rinse the internal canal of the insert with
distilled water injected under pressure
(3,8 bar) for at least 10 seconds to
eliminate all residues;
14
3,8 bar
10 sec.
4.2.3.5 Inserts Torque Wrench
Put the torque wrench to soak in the
enzymatic detergent solution
h)
for 10
minutes at 40 °C ±2 °C;
h) Process validated by independent body with
enzymatic detergent Enzymec (Mectron).
15
During immersion in the enzymatic
solution, delicately brush the surface of
the torque wrench with a brush with soft
nylon bristles to eliminate all traces of
visible dirt, both in the internal part and
the external one;
16
Delicately brush the surface of the wrench
with a brush with soft nylon bristles under
tap water;
17