CLEANING AND STERILIZATION
Rinse the torque wrench with distilled
water until it is clearly visible that the
contaminants have all been removed;
18
Place the torque wrench in the ultrasonic
tank submerged by the enzymatic
detergent solution at 40 °C ±2 °C, for at
least 10 minutes;
19
Rinse with tap water to eliminate all
residues;
20
Delicately brush the surface of the torque
wrench with a brush with soft nylon
bristles using distilled water;