CLEANING AND STERILIZATION
4.2.2.4 Inserts
Place the insert horizontally in a clean
container and add enzymatic detergent
solution
d)
until the insert is covered
completely;
d) Process validated by independent body with
enzymatic detergent Enzymec (Mectron).
5
Let the insert soak for 10 minutes at 40
°C ±2 °C. This procedure reduces the
amount of blood, protein and mucous
present on the insert;
6
Using a 20 ml disposable syringe, inject
20 ml of enzymatic detergent solution
3 times inside the cavity of the insert to
ecaciously remove the residues from
the internal surface;
7
x3
Delicately brush the surface of the insert
under running water (drinking water,
temperature 20 °C +/-2 °C), in particular
the parts that are dicult to access
(cavities, structured surfaces, etc.) using a
brush with soft nylon bristles;
The insert must be brushed until no
residues can be seen (at least 30 seconds);
8