4.1 Cleaning of the Console and of the Foot Pedal
4.1.1 Preparation
Make sure that all the following accessories
have been removed/disconnected from the
device console (see
Chapter 3 on page 5
):
electrical power supply cable;
foot pedal;
peristaltic pump tube;
handpiece;
handpiece xed support;
irrigation bag support rod;
protective foil (if present);
supplementary grounding connector
(if present).
WARNING: Do not sterilize the device
console and the foot pedal: it could stop
working and cause harm to person and/or
damages to things.
WARNING: Always turn the device o
via the switch and disconnect it from the
electrical network before performing the
cleaning and disinfection interventions.
WARNING: The device is not protected
against penetration of liquids. Do not spray
liquids directly onto the surface of the device.
CAUTION: Do not clean the device
console and the foot pedal under running tap
water.
CAUTION: Do not immerse the device
console and the foot pedal in liquids and/or
solutions of various nature.
4.1.2 Materials Required
Clean, soft cloths with low ber-release;
Detergent solution (pH 6-9).
CAUTION: If you wish to disinfect, we
recommend that you use water-based
disinfectant solutions with neutral pH (pH7).
Alcohol-based disinfectant solutions and
hydrogen peroxide are contra-indicated,
because they can fade the color and/or
damage the plastic materials. This also holds
true for chemical products such as acetone
and alcohol.