20
4.2.3 Manual Cleaning
4.2.3.1 Materials Required
Enzymatic detergent with pH 6-9;
Water;
Container for the immersion in the
enzymatic solution;
Ultrasound tank;
Clean, soft cloths with low ber-release;
Brush with soft nylon bristles;
Syringe;
Demineralised water.
CAUTION: If you wish to disinfect, we
recommend that you use water-based
disinfectant solutions with a neutral pH
(pH7). Alcohol-based disinfectant solutions
and hydrogen peroxide are contra-indicated,
because they can fade the color and/or
damage the plastic materials. This also holds
true for chemical products such as acetone
and alcohol. Always rinse with sterile water to
preserve the disinfection.
CAUTION: Do not use hydrogen
peroxide. If you wish to disinfect the
accessories, do not use hydrogen peroxide,
but only disinfectants with neutral pH (pH7);
always rinse with sterile water to preserve the
disinfection.
4.2.3.2 Preparation
Prepare an enzymatic detergent solution
with pH 6-9
f)
, as per the instructions of
the manufacturer;
CAUTION: Dispose of the enzymatic
detergent correctly once used. Do not
recycle.
f) Process validated by independent body with
enzymatic detergent Enzymec (Mectron).
1
4.2.3.3 Handpiece with Cord
Clean the surface of the handpiece and of
its cord with a clean, soft cloth with low
ber-release, dampened with a detergent
solution (pH 6-9);
2