26
4.2.3.6 Peristaltic Pump Tubes and Mobile Handpiece Support
Immerse the peristaltic pump tube and
the handpiece mobile support in the
enzymatic detergent solution prepared;
22
Rinse with abundant distilled water;
23
4.2.3.7 Tube-to-Tube Connector
Place the connector in the ultrasonic tank
submerged by the enzymatic detergent
solution at 40 °C ±2 °C, for at least 10
minutes;
24
Remove the connector from the
ultrasonic tank, brush it with the brush
with soft nylon bristles, and rinse it with
distilled water to eliminate all residues of
detergent;
25