30
CAUTION: Before starting the
sterilization cycle, make sure that the
insert is thoroughly dry, both externally
and internally. For this purpose, blow
compressed air both externally and
through the internal thru hole; this will
prevent the onset of stains and haloes
on the surface or rust inside the insert.
2
Thoroughly dry the accessories on all
their surfaces, internal and external, using
a clean and soft cloth with low ber-
release or by blowing compressed air,
depending on the case.
3
Dry the torque wrench using a clean and
soft cloth with low ber-release;
4
Lubricate with medical degree lubricant
in the indicated point;
CAUTION: Do not use lubricants
based on oil or silicon.
5