45
Specification
Circuit test
Test Voltage: 12 V p - p, 50Hz square wave. (Typical)
Max current: 100 mA (typical)
Indications: Open circuit, OK, Short circuit.
Insulation test
Range: 0 - 50 MΩ
Accuracy: 5.0% ± 100 kΩ
Compliance: 500 V at 0.5 MΩ
Pass Band options: 0.25MΩ, 0.3MΩ, 0.5MΩ, 1MΩ, 2MΩ, 4MΩ, 5MΩ, 7MΩ, 8MΩ, 10MΩ, 20MΩ, 50MΩ
Earth Bond Test
Range: 0 - 999 mΩ Resolution 1 mΩ
1000 – 1999mΩ Resolution 10 mΩ
Accuracy: 0 <R> 499 mΩ 5.0% ±5mΩ
500 m <R> 1999 mΩ 5.0% ±50mΩ
Test Voltage: 10 V a.c. (Typical)
Test Current: 10A or 25A into 0.1 (typical)
Pass Bands: 50 mΩ, 100 mΩ, 200 mΩ, 300 mΩ, 400mΩ, 500mΩ, 600mΩ, 750mΩ, 800mΩ, 1Ω,
1.5, 2
Earth Continuity Test
Range: 0 - 999 mΩ Resolution 1 mΩ
1.00 - 9.99Ω Resolution 10 mΩ
Accuracy: 5.0% ±10mΩ
Test Voltage: 4 V d.c. (Typical)
Test Current: 200 mA short circuit (typical)
Pass Bands: 50 mΩ, 100 mΩ, 200 mΩ, 500 mΩ, 750mΩ, 1Ω, 1.5Ω, 2Ω, 2.5Ω, 5Ω, 7.5Ω, 10Ω
Operation Test
Range: 0 - 3.0 kVA.
Accuracy: 0 <VA> 99 VA: 5.0% ± 5 VA
100 <VA> 999 VA: 5.0% ± 10 VA
1.0 <VA> 3.0kVA: 5.0% ± 100 VA
Reference: Reading corrected to 230 V
Pass Band options: 50 VA, 100 VA, 200 VA, 350 VA, 500 VA, 700 VA, 1000 VA, 1200 VA, 1500VA,
2000 VA, 2500 VA, 3000 VA.