46
Specification
Earth Leakage Test
Range: 0 - 15 mA
Accuracy: ±5.0% ± 100 µA (@ 50Hz) (±20% @ 60Hz)
Reference: Reading corrected to 253 V
Pass Band options: 0.25mA, 0.5mA, 0.75mA, 1mA, 1.5mA, 2mA, 3.5mA, 5mA, 7mA, 7.5mA, 10mA, 15mA
Flash Test (PAT4 DVF)
Range: 0 – 3.5mA
Accuracy: 5 % 100 µA
Test voltage: Class 1 – 1500 V a.c. (typical)
Class 2 – 3000 V a.c. (typical)
Pass Band Options: 0.2mA, 0.3mA, 0.4mA, 0.5mA, 0.6mA, 0.75mA, 1mA, 1.2mA, 1.5mA, 2mA, 2.5mA, 3mA
Fuse Test
Test Voltage: 5 V d.c. typical
Test Current: 500 µA short circuit typical
Indication: Audible using internal bleeper
Extension Lead test
Test Voltage: 5V d.c. typical
Test current: 1 mA short circuit typical
Indication: Lead short circuit (Live to Neutral)
Lead open circuit (Live or Neutral)
Live / Neutral transposed
Lead OK
Bond and Insulation test follow OK condition.
Mains supply
Input Voltage Range: The PAT4 DV contains an automatic mains voltage changeover system. To test 110V appliances an external 110 Volts source must be provided.
Operating Voltage: 230 Volt: 207 - 253 Volts 50Hz / 60Hz*
110 Volt: 99 – 121 Volts 50Hz.
Fusing: FS 13 Amp to BS1362.
Safety Specifications
Meets the requirements of EN 61010-1 phase to earth Installation Category II** for Safety Class 1
EMC: In accordance with IEC61326-1
Operational uncertainties: Refer to www.megger.com
* See Earth Leakage tolerance for 60Hz operation.