Chapter 2 – Device Description
15
The display switches to the initial state after every switch-on, where the
current time and the chamber pressure in bar and the (steam) temperature
in °C are displayed.
The steam sterilizer requires
feed water
for steam production according to
DIN EN 13060, Appendix C. The feed
water supply is provided either by an external feed water container or with
a water treatment unit (e.g. MELA
Feed water via the external
If you employ the feed water container for the feed water supply, you must
from time to time fill it manually.
At the given point in time, the
steam sterilizer outputs an appropriate
The external feed water container holds 5 litr
es and suffices for approx.
20
sterilizations depending on the selected program.
demand for a program run is 280 ml.
A water treatment unit is connected to the domestic water supply. It
produces the
→feed water required for generating the steam for the
If you use a water treatment unit, then there is always sufficient feed water
available. You no longer have to manually fill the feed water container.