Chapter 3 – Initial Commissioning
17
B= 43 cm
C=47 cm
D= 29 cm
E= 9 cm
F= 5 cm
G= 10 cm
H= 22 cm
B= 43 cm
C=46 cm
D= 29 cm
E= 9 cm
F= 5 cm
G= 10 cm
H= 22 cm
Additional space for feed
water supply
Other than for the steam sterilizer, the feed water and the waste water
container require space.
Space requirement for water
treatment unit
The attachment can be next to the steam sterilizer or above or below
the steam sterilizer
Attachment next to the steam sterilizer or above or below the steam
sterilizer (e.g. lower cabinet)
MELAdem 47 module housing
MELAdem 47 water storage tank
The feed water and waste water container should be installed in
immediate proximity, below the
steam sterilizer and easily accessible.
The length of the connection hoses
and the cables amounts to 1.5 m.
During installation they can be cut to the desired l
and cables up to a maximum of 5 m may be ordered and used.
The mains socket must be freely accessible after installation so that the
device can be taken from the electricity supply at any time.
The feed water connection hose (Ø 6 mm) is inserted to the end into the
feed water connector (
Fig. 1/(14)) on the rear side of the steam sterilizer
cured with the union nut. In order to enquiry the water level in the
container, the cable is connected on the
rear side of the steam sterilizer
The waste water connection hose (Ø 8 mm) is inserted to the end into the
waste water connector (
Fig. 1/(11)) on the rear side of the steam sterilizer
and secured with the union nut. In order to enquiry the water level in the