Chapter 4 – Sterilization
29
For difficult drying tasks you can extend the drying time of a program by
50% with the function Additional drying.
For this proceed as follows:
When starting the program, simultaneously press the KEY (S)
AND KEY (+).
Additional drying selected. Then the program
If you selected a program with the program selection button (P), in addition
to the selected program, the sterilization temperature is displayed and
whether the respective program is
suitable for wrapped or unwrapped
(S), in order to start the program.
The steam sterilizer checks the transport of the
-Program S is started, the warning text Warning only
If the load exclusively contains unwrapped instruments, press the KEY
(S) once again to confirm and start the program.
After starting the program, you can follow the program run in the
display.
Chamber temperature and pressure as well as the remaining duration until
end of sterilization respectively the expired drying time is displayed.
In the evacuation phase the
fractionated flow method with pulsating,
repeated steam entry and discharge, sucks the air out of the chamber and
introduces saturated steam into the chamber.
Depending on selected program and the current chamber temperature at
program start, additional
The heating-up phase follows after the evacuation phase. A pressure
discharge with a corresponding display is carried out at the end of the
sterilization phase.
After achieving the sterilization parameters of pressure and temperature,
the sterilization phase begins.
In alternation with pressure and temperature the remaining time of the
sterilization phase is also displayed.
A pressure discharge with a corresponding display is carried out at the end
of the sterilization phase.
The drying phase begins after the pressure discharge.
The regular drying time is approx. 1
minute for the Quick-Program S and
approx. 10 minutes for all other programs.
Additional drying
selected
Universal-Program S
134℃ wrapped
1. Fractionation
0.69 bar 115°C
Sterilization
2 min., 12 s left
Current-drying
since 1’ 0.9 bar 85 °C